의료용 X-선 조사 시 개인피폭선량계의 방향 및 에너지 의존성에 관한 연구

Metadata Downloads
Author(s)
김호성
Issued Date
2013
Abstract
Purposes
We measured the scattered radiation occurred when low, medium, high energy x-ray were radiated to a human body phantom respectively by using 3 types of personal dosimeter; TLD, PLD, OSLD under the clinical environment.
So that, we intended to understand the characteristics of the scattered radiation with different distances and directions and select the best dosimeter that has the lowest energy and direction dependency to present the basic information which can be the standards for the new personal dosimeter hereafter.
Methods
An anthropomorphic phantom was irradiated by a general x-ray machine for low energy, a fluoroscopy for medium energy while for high energy is a linear accelerator of radiation therapy machine and scattered radiation was measured by an ion chamber to check traceability of the standard source.
We used 3 types of personal dosimeters including TLD, PLD, OSLD to test the sensitivity according to the direction and distance. For each experiment, we repeated the measurements more than 10 times to improve accuracy and comparatively analyzed the angular dependence by using the one way ANOVA method to understand differences on the distance and angle. And also, if there are any considerable differences between the groups from ANOVA methods, we applied SCHEFF’S method for further cause analysis.
Results
In the low energy category, the OSLD was best in sensitivity depending on distance, amount of radiation and intensity. The TLD is most accurate with kVp variance in energy dependency. The PLD is most superior with mAs variance in energy dependency. There are no differences in angular dependencies.
In the medium energy category, the TLD is most accurate in sensitivity and the PLD is most accurate in angular dependency.
In the high energy category, the PLD is most accurate from all of the dependencies.
Conclusion
These three devices are universally used as individual radiation monitoring devices. Results of the three measuring devices’ indicate that the OSLD is most accurate in determining low energies and the PLD is most accurate in determining the medium and high energy levels. And also the PLD is most accurate in angular dependencies in whole energy level.
Based on the results that were acquired in a clinical environment, it is important to select the appropriate monitoring device that is best suited for the energy level and type of equipment being used. This will ensure that the safety of health care workers is maintained.
Alternative Title
A study on the direction and energy dependency of personal exposure dosimeter for medical x-ray irradiation
Alternative Author(s)
kim, Ho Sung
Affiliation
조선대학교 원자력공학과 대학원
Department
일반대학원 원자력공학과
Advisor
정운관
Awarded Date
2013-02
Table Of Contents
목 차

표 목차 ⅱ
그림 목차 ⅳ
ABSTRACT vi
I. 연구배경 1
II. 연구목적 4
III. 실험기기 및 실험방법
A. X-선 발생장치의 에너지 영역과 영역별 실험기기 4
B. 실험방법 11
IV. 결과
A. 저 에너지 영역 17
B. 중 에너지 영역 37
C. 고 에너지 영역 46
V. 고찰
A. 선량계의 측정원리 및 구조 64
B. 선량계의 판독 76
C. 선형가속기 79
VI. 결론 91
참고문헌 96
Degree
Doctor
Publisher
조선대학교 대학원
Citation
김호성. (2013). 의료용 X-선 조사 시 개인피폭선량계의 방향 및 에너지 의존성에 관한 연구.
Type
Dissertation
URI
https://oak.chosun.ac.kr/handle/2020.oak/9832
http://chosun.dcollection.net/common/orgView/200000263754
Appears in Collections:
General Graduate School > 4. Theses(Ph.D)
Authorize & License
  • AuthorizeOpen
  • Embargo2013-01-02
Files in This Item:

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.