치과용 임플란트 고정체 표면의 형상이 부식특성에 미치는 영향

Metadata Downloads
Author(s)
박진태
Issued Date
2008
Abstract
This study was performed to investigate the effect of surface morphology on the electrochemical characteristics of dental implant fixture. Titanium implant was selected with sort of 5 types made by domestic company. It has different surface treatment and structure of screw, respectively.
Estimation of electrochemical stability of dental implant with various types in 1M H3PO4 electrolytes have been researched by using electrochemical methods. Specimen was used for corrosion test in 0.9 % NaCl solution.
The surface morphology were observed using scanning electron microscopy (SEM) and energy dispersive x-ray spectroscopy(EDX). The electrochemical stability was investigated using potentiostat(EG&G Co, 263A). The corrosion surface were observed using scanning electron microscopy(SEM).

The results were as follows:
1.From the results of potentiodynamic test in 0.9% NaCl solution, the current density from sample 1,2,3 and 5 of implant fixture with RBM treatment tested in 0.9% NaCl was lower than that from sample 4 of implant fixture tested in 0.9% NaCl solution. Especially, sample 4 of implant fixture with SLA treatment was presented the lowest value of corrosion resistance.
2.From the results of SEM micrographs, sample 1, 2, 3, 5 of RBM treatment were observed rough surface with sharp of low ridge and valley, sample 4 of SLA treatment was observed irregular rough surface with sharp as a needle.
3.From the A. C. impedance test results, polarization resistance(Rp) value of sample 4, were lower than those of sample 1, 2, 3, 5.
4.In case of SLA treated implant fixture, the pit and corrosion trace were observed in valley and crest of implant surface.
Alternative Title
Effect of Surface Morphology on the Electrochemical Characteristics of Dental Implant Fixture
Alternative Author(s)
Park, Jin Tae
Affiliation
일반대학원 치의학과
Department
일반대학원 치의학과
Advisor
고영무
Awarded Date
2009-02
Table Of Contents
영문초록 ⅴ

1. 서 론 1

2. 실험재료 및 방법 3
2-1. 시편준비 3
2-2. 전기화학적 시험평가 3
2-2-1. 동전위 분극시험 4
2-2-2. 교류 임피던스 시험 5

3. 실험 결과 및 고찰 6
3-1. 주사전자현미경에 의한 임플란트 표면 관찰 6
3-2. 동전위 분극시험 14
3-3. 교류 임피던스 시험 16

4. 결 론 20

참고문헌 21

감사의 글
Degree
Master
Publisher
조선대학교
Citation
박진태. (2008). 치과용 임플란트 고정체 표면의 형상이 부식특성에 미치는 영향.
Type
Dissertation
URI
https://oak.chosun.ac.kr/handle/2020.oak/8115
http://chosun.dcollection.net/common/orgView/200000237850
Appears in Collections:
General Graduate School > 3. Theses(Master)
Authorize & License
  • AuthorizeOpen
  • Embargo2009-02-04
Files in This Item:

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.