The purpose of this study was to evaluate the possibility of using digital model and CBCT image for model analysis. CBCT images, plaster models, digital models of 20 orthodontic patients with permanent dentition were used for comparison of model analysis measurements (tooth size, arch width, required space, available space, arch length discrepancy, Bolton tooth size discrepancy)
The results were as follows.
1. Absolute deferences of repeated measurements of tooth size and dental arch width were 0.06 to 0.27 mm, 0.26 to 0.84 mm in plaster models, 0.01 to 0.20 mm, 0.25 to 0.63 mm in digital models and 0.15 to 0.26 mm, 0.34 to 0.49 mm in CBCT images.
2. The average differences of tooth size measurements between digital model and plaster model were 0.01 to 0.20 mm, and digital model showed significant differences in the 7 tooth sizes.
3. The average differences of tooth size measurements between CBCT image and plaster model were -0.22 to 0.01 mm, and CBCT image showed significant differences in the 5 tooth sizes.
4. The average differences of arch length discrepancy measurements between digital model and plaster model were below 0.82 mm, and the difference of Bolton discrepancy measurements was below 0.44 mm. There was no statistically significant difference.
5. The average differences of arch length discrepancy measurements between CBCT image and plaster model were below 0.43 mm, and there was no statistically significant difference. Overall Bolton discrepancy measurements was 1.25 mm smaller in the CBCT images and was statistically significant.
In conclusion, although there were statistically significant differences in some model analysis measurements between digital model and plaster model, and between CBCT image and plaster model, the ranges of measurement errors were clinically acceptable. Therefore, digital model and CBCT image can be used for model analysis measurements.
Alternative Title
Comparison of model analysis among plaster model, digital model using laser scan, and cone beam CT image
Alternative Author(s)
Lim, Mi-Young
Affiliation
조선대학교 대학원
Department
일반대학원 치의공학과
Advisor
임성훈
Awarded Date
2008-08
Table Of Contents
I. 서론 = 1
II. 연구대상 및 방법 = 4
1. 연구대상 = 4
2. 연구방법 = 4
3. 통계처리 = 6
Ⅲ. 연구 성적 = 7
Ⅳ. 총괄 및 고안 = 18
Ⅴ. 결론 = 27
참고문헌 = 29
Degree
Doctor
Publisher
조선대학교 대학원
Citation
임미영. (2008). 석고 모형, 레이저 스캔 방식 디지털 모형, 콘 빔 CT 영상 간의 모형 분석 비교.
Type
Dissertation
URI
https://oak.chosun.ac.kr/handle/2020.oak/7329
http://chosun.dcollection.net/common/orgView/200000236596
Appears in Collections:
General Graduate School > 4. Theses(Ph.D)
Authorize & License
  • AuthorizeOpen
  • Embargo2008-07-18
Files in This Item:

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.