CHOSUN

TiN 코팅된 지대주 나사의 마모에 관한 연구

Metadata Downloads
Author(s)
정석원
Issued Date
2007
Keyword
TiN|코팅|지대주 나사|마모|Screw
Abstract
The purpose of this study was to evaluate surface changes and abrasion of the implant fixture and TiN coated abutment screw after repeated delivery and removal under SEM and universal measuring microscope. Implant systems used for this study were Osstem and 3i. Seven pairs of implant fixture, abutment and abutment screws for each system were selected and all fixtures were perpendicularly mounted in liquid unsaturated polyesther with dental surveyor. Titanium abutment screws of Osstem and 3i were assigned to Group 1 and Group 2 respectively. TiN coated Abutment screws of Osstem and 3i were classified to Group 3 and Group 4. After 20 times of repeated closing and opening test, the evaluation for the change of inner surface of implant and TiN-coated abutment screw, and measurement of weight loss were performed. Mann-Whitney test with SPSS statistical software for Window was applied to analysis the measurement of weight loss.
The results were as follows ;
1. Mean weight loss of abutment screws for four groups was 0.181± 0.145mg in Group 1, 0.069±0.023mg in Group 2, 0.056± 0.026mg in Group 3 and 0.043±0.016mg in Group 4.
2. In the comparison of weight loss of Group 1 and Group 3, there was a statistically significant difference between Group 1 and Group 3 (P<0.05, Mann-Whitney test).
3. In the comparison of weight loss of Group 2 and Group 4, Group 2 revealed a statistically higher mean value than Group 4(P<0.05, Mann-Whitney test).
4. Abrasion and defect of mating thread surface of abutment screw were considerable in both Group 1 and Group 2, while only slight abrasion and defect were founded in Group 3 and Group 4.
5. In the SEM view of internal mating surface of implant, Group 1 and Group 2 showed a heap of titanium particles formed by galling and seizing tendency of titanium on the mating surface while Group 3 and Group 4 revealed slight abrasion or scratches on the mating surface.
Conclusively, TiN coating of abutment screw showed better resistance to abrasion than titanium abutment screw. It was concluded that TiN coating of abutment screw would reduce the loss of preload with good abrasion resistance and low coefficient of friction, and help to maintain screw joint stability.
Alternative Title
Abrasion of TiN coated Abutment Screw
Alternative Author(s)
Jeong, Seok-Ueon
Affiliation
조선대학교 대학원
Department
일반대학원 치의학과
Advisor
정재헌
Awarded Date
2007-08
Table Of Contents
Ⅰ. 서론 = 1
Ⅱ. 연구 재료 및 방법 = 3
Ⅲ. 연구 성적 = 9
Ⅳ. 총괄 및 고안 = 14
Ⅴ. 결론 = 18
References = 20
Degree
Doctor
Publisher
조선대학교 대학원
Citation
정석원. (2007). TiN 코팅된 지대주 나사의 마모에 관한 연구.
Type
Dissertation
URI
https://oak.chosun.ac.kr/handle/2020.oak/6838
http://chosun.dcollection.net/common/orgView/200000234326
Appears in Collections:
General Graduate School > 4. Theses(Ph.D)
Authorize & License
  • AuthorizeOpen
  • Embargo2007-11-13
Files in This Item:

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.