완속기시(Soft-start) 광조사 방식이 복합레진의 중합 및 수축응력에 미치는 효과

Metadata Downloads
Author(s)
위유민
Issued Date
2004
Abstract
The purpose of this study was to evaluate the influence of soft-start light curing on contraction stress and hardness of composite resin.
Composite resin in acrylic mold (inner diameter 7mm, outer diameter 10mm, height 2mm) was cured using the one step continuous curing method with three different light sources; conventional halogen light curing for 40 seconds at 400 mw/㎠, plasma arc light curing for 6 seconds at 1300 mW/㎠ and LED light curing for 10 seconds at 700 mW/㎠. And the soft-start curing modes with two different light sources; plasma light curing for 2 seconds light exposure at 650 mW/㎠ followed by 3 seconds at 1300 mW/㎠ and LED light curing with exponential increase for 5 seconds followed by 10 seconds at 700 mW/㎠ were used.
Contraction stress was measured using strain gauge method and also Vickers hardness was measured 24 hours after polymerization at the top and bottom of specimens.
Resin-acrylic interfaces were observed using a scanning electron microscope(SEM).
The results of present study can be summarized as follows:
1. Composite resin in acrylic molds showed the initial expansion at the early phase of polymerization.
2.Contraction stresses at 10 min after polymerization were significantly reduced with the soft-start curing both in plasma and LED light sources(P<0.05).
3.Plasma light curing with soft-start resulted in not only the lowest contraction stress, but also the lowest hardness (P<0.05).
4. LED light curing with soft-start showed lower contraction stress than the one-step continuous halogen and LED light curing (P<0.05).
5.Microhardness of specimens cured by LED light with soft-start was equivalent to that of cured by the one-step continuous halogen and LED light (P>0.05).
6.Curing by LED light with soft-start and conventional halogen light resulted in better marginal sealing than plasma light and one-step LED light curing.
On the basis above results, we think that LED light curing with soft-start provided the reduction of polymerization contraction stress, the adequate hardness of surfaces and the proper marginal sealing.
Alternative Title
Effect of soft-start light curing on the polymerization and the contraction stress of composite resin
Alternative Author(s)
Wee, You-Min
Affiliation
조선대학교 대학원
Department
일반대학원 치의학과
Advisor
이상호
Awarded Date
2005-02
Table Of Contents
목차
Abstract = ⅳ
Ⅰ. 서론 = 1
Ⅱ. 실험재료 및 방법 = 4
Ⅲ. 실험 성적 = 10
Ⅳ. 총괄 및 고안 = 16
Ⅴ. 결론 = 23
참고문헌 = 25
사진부도 설명 = 32
사진부도 = 34
Degree
Doctor
Publisher
조선대학교 대학원
Citation
위유민. (2004). 완속기시(Soft-start) 광조사 방식이 복합레진의 중합 및 수축응력에 미치는 효과.
Type
Dissertation
URI
https://oak.chosun.ac.kr/handle/2020.oak/5699
Appears in Collections:
General Graduate School > 4. Theses(Ph.D)
Authorize & License
  • AuthorizeOpen
  • Embargo2005-10-13
Files in This Item:
  • There are no files associated with this item.

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.